Single-chamber compact implantable telemetry pacemakers of SSIR VIRSAR SR and SSI VIRSAR SC types are designed for treatment of heart rhythm disorders by electrical stimulation of ventricles or atria.
Brand
Baikal
Baikal

Single-chamber pacemakers

Single-chamber compact implantable telemetry pacemakers of SSIR VIRSAR SR and SSI VIRSAR SC types are designed for treatment of heart rhythm disorders by electrical stimulation of ventricles or atria.
Brand
Baikal
Baikal

Description

About the production facility

Pacemakers are produced at the Izhevsk Mechanical Plant JSC. The company has extensive experience in the design and serial production of implantable pacemakers, programmers and bipolar electrodes. With its modern high-tech equipment and highly competent personnel, the plant boasts over 30 years of experience in the manufacture of medical appliances. 

Single-chamber pacemakers

Single-chamber compact implantable telemetry pacemakers of SSIR VIRSAR SR and SSI VIRSAR SC types are designed for treatment of heart rhythm disorders by electrical stimulation of ventricles or atria.

The VIRSAR SR pacemaker has a frequency adaptation feature. Its activity sensor allows for increasing the stimulation pulse frequency depending on the level of patient's physical activity.

The pacemakers can be used with unipolar or bipolar electrodes with IS-1 standard leads (3.2 mm). The pacemakers use screwless connections to attach the anode contact of the connector part of the electrode.

The pacemakers can be programmed with the UNI-2 programmer. 

Features

  • extensive statistics on multiple events presented as convenient histograms,
  • sleep rate and intracardiac electrogram recording features,
  • temporal programming for rates from 30 to 460 ppm,
  • electrode impedance measurement from 200 to 3,000 ohms,
  • emergency stimulation mode with 6.3 V amplitude and stimulation pulse duration of 976 μs.

Characteristics

VIRSAR SR pacing modes
VVI, VVIR, VVT, VVTR, VОО, VOOR, AAI, AAIR, AAT, AATR, AОО, AOOR
VIRSAR SC pacing modes
VVI, VVT, VОО, AAI, AAT, AОО
Pacing pulse rate, ppm
30...60...160, 5 ppm increments
Pacing pulse amplitude, V
0.60; 1.00; 1.35; 1.75; 2.10; 2.50; 2.90; 3.25; 3.65; 4.00; 4.40; 4.80; 5.15; 5.55; 5.90; 6.30; 7.50; 9.00
Pacing pulse duration, μs
122...366...976, 122 µs increments
P-wave sensitivity threshold, mV
0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0
R-wave sensitivity threshold, mV
1.0; 2.0; 3.0; 3.0; 4.0; 5.0; 6.0; 7.0; 8.0
Refractory period, ms
200...260...500, 20 ms increments
Rate hysteresis, ms
OFF; 50...300 s, 50 ms increments
Rate hysteresis, ms
OFF; 50...300 s, 50 ms increments
Rate search hysteresis, ms
OFF; ON
Sleep rate, ppm
OFF; -5...-30, 5 ppm increments
Bedtime, h 0...23, 1 h
increments; 0...50 min, 10 min increments
Waking time, h
0...23, 1 h increments; 0...50 min, 10 min increments
VARIO mode programming by amplitude and duration
OFF; ON
Polar input-output mode
unipolar; bipolar
Telemetry
Model and serial number, programmed pacing parameters, battery status, electrode impedance, event markers, patient data
Pacemaker weight, g
23
Pacemaker dimensions, LxWxH, mm
51.5x43x7.6
Power source
lithium/fluorocarbon cell, 1.0 Ah
VIRSAR SR rate adaptation modes
Maximum sensor rate, ppm
90...180 s, 5 ppm increments
Reaction time, s
10...50 s, 10 s increments
Recovery time, min
1...16, 1 min increments
Rate rise coefficient
AUTO; 1...16, 1 pt increments